EN-17
Operation
Note: Before operating the 1488 HD Camera, ensure all system components
have been set up according to the instructions in the “Setup and
Interconnection section.
Operating the Camera with a Light Source
IMPORTANT SAFETY NOTICE: Before operating this device,
please read this operating manual thoroughly and carefully.
When using this device with a light source, re and/or severe
injury may result to the patient, user or inanimate objects if the
instructions in this manual are not followed.
All light sources can generate signicant amounts of heat
(exceeding 41°C/106°F) at the scope tip, the scope light post,
the light cable tip, and/or near the light cable adapter. Higher
levels of brightness from the light source result in higher levels
of heat. Always adjust the brightness level of the camera and the
monitor before adjusting the brightness level of the light source.
Adjust the brightness level of the light source to the minimum
brightness necessary to adequately illuminate the surgical site.
In addition, adjust the internal shutter of the camera higher in
order to run the light source at a lower intensity. Avoid touching
the scope tip or the light cable tip to the patient, and never
place them on top of the patient, as doing so may result in burns
to the patient or user. In addition, never place the scope tip, the
scope light post, the light cable adapter, or the light cable tip
on the surgical drapes or other ammable material, as doing so
may result in re.
Always place the light source in standby mode whenever
the scope is removed from the light cable or the device is
unattended. The scope tip, scope light post, light cable adapter,
and light cable tip will take several minutes to cool o after
being placed in standby mode, and therefore may still result in
re or burns to the patient, user, or inanimate objects.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker 1488 and is the answer not in the manual?

Stryker 1488 Specifications

General IconGeneral
Resolution1920 x 1080
Sensor TypeCMOS
White BalanceAutomatic/Manual
OutputHDMI, DVI, HD-SDI
Camera Head TechnologyThree-chip

Related product manuals