Period of additional logging after triggering
condition fulfilment.
Recording of the waveform signal in the wave file.
Switching on the signal recording: Continuous or
on trigger: Slope +, Slope -, Level +, Level -,
Gradient +, Trigger manual or Integr. Period.
Continuous means that the waveform signal will
be recorded continuously from the start of the
measurement till its end. On trigger recording put
additional conditions for triggering and ending of
the recording.
If the instrument is waiting for the trigger condition,
the appropriate trigger icon is flashing on the
display alternatively with the „signal” icon.
Type of trigger that starts the signal recording for
Recording Time under the condition: rising value
of the LEQ result (Source) integrated during 0,5
ms passes above the threshold level (Level).
Type of trigger that starts the signal recording for
Recording Time under the condition: falling value
of the LEQ result (Source) integrated during 0,5
ms passes below the threshold level (Level).
Type of trigger that starts the signal recording for
Recording Time under the condition: value of the
LEQ result (Source) integrated by the 0,5 ms
period is greater than the threshold level (Level).
Type of trigger that starts the signal recording to
start for Recording Time under the condition:
value of the LEQ result (Source) integrated by the
0,5 ms period is lower than the threshold level
(Level).