SVAN 973 User Manual
93
Gradient +
Type of trigger that starts the signal recording for
Recording Time under the condition: value of the
LEQ result (Source) integrated by the 0,5 ms
period is greater than the threshold level (Level)
and the gradient of this Source is greater than the
threshold level (Gradient).
Manual
4.5.4
Trigger manual
Type of trigger that starts manual triggering of the
signal recording start after pressing
simultaneously and keys during the
measurement.
Manual
4.5.4
Integr. Period
Type of trigger that starts the signal recording for
Recording Time every time the measurement
starts. If Integration Period is shorter than
Recording Time, the event recording will be
continued for additional Recording Time.
Manual
4.5.4
Filter
Weighting filter used during signal recording: Z, A,
C, B or LF.
Manual
4.5.4
Sampling
Sampling frequency of the event recording:
24 kHz or 12 kHz.
Manual
4.5.4
Signal Gain
Gain of the recorded signal: 0 dB 40 dB.
Manual
4.5.4
Source
Measured result that is compared with the
threshold level for triggering (Level) LEQ
measured in the first profile: Leq(1).
Manual
4.5.4
Level
Threshold level of the Source for the trigger
condition fulfilment.
Manual
4.5.4