Time interval of checking the triggering conditions.
This parameter can be set as: Logger Step,
0.5 ms, 100.0 ms and 1 s.
Threshold level of the source signal value speed
of changing (Gradient) for trigger condition
fulfilment.
Period of signal recording before the first trigger
condition moment: Off or 1 s.
Time of the signal recording after meeting every
trigger condition. The available values can be
selected from 1 s to 8 h. Recording stops after
Recording Time or earlier if the measurement is
stopped manually.
Screen that enables setting the 1/1 Octave or
1/3 Octave spectrum parameters setup: Filter and
Detector.
Position in the Logger Results screen that
switches on/off the Lpeak spectra saving as a
time-history in a logger file.
Position in the Logger Results screen that
switches on/off the Leq spectra saving as a time-
history in a logger file.
Weighting filters for the 1/1 Octave and
1/3 Octave analysis: A, B, C, Z.