Chapter 5 Command Reference
Description of SCPI Commands
82 Operating and Programming Manual
DIAGnostic :VOLTage:COVen?
This query returns the Cesium oven heater voltage.
Response format = <NR3>, precision = 0.1 V.
DIAGnostic :VOLTage :EMULtiplier?
:This query returns the electron multiplier voltage setpoint.
Response format = <NR3>, precision = 1 V.
DIAGnostic :VOLTage:HWIonizer?
This query returns the hot wire ionizer voltage setpoint.
Response format = <NR3>, precision = 1 x 10
-1
V.
DIAGnostic :VOLTage:MSPec?
This query returns the mass spectrometer voltage setpoint.
Response format = <NR3>, precision = 1 x 10
-1
V.
DIAGnostic :VOLTage:PLLoop?
This query returns tuning voltages for each of four phase locked loops: 9.2 GHz,
640 MHz, 87 MHz, 9 MHz.
Response format = <NR3, NR3, NR3, NR3>, precision = 1 x 10
-1
V.
DIAGnostic :VOLTage:ROSCillator?
This query is equivalent to ―[SOURce:]ROSCillator:MVOLtage?‖ See
[SOURce:]ROSCillator:MVOLtage?
DIAGnostic :VOLTage:SUPPly?
This query returns voltages from the +5, +12, and -12 volt power supplies.
Response format = <NR3, NR3, NR3>, precision = 1 x 10
-1
V.