Chapter 5 Command Reference
Description of SCPI Commands
Operating and Programming Manual 89
ROSCillator:STEer? [MIN|MAX]
This query returns the active value of fractional frequency offset or minimum-
maximum acceptable range limits.
Response format = <NR3>, precision = 1 x 10
-15
.
ROSC:FREQuency[1|2] <frequency>
This command sets the output frequency of the indicated port with the supplied
parameter.
The 5071A must be in REMote mode to execute this command.
The frequency parameter is 5 x 10
6
or 10 x 10
6
Hz.
Values are rounded to the nearest valid setting, unless out of range.
The function of this command is not affected by *RST.
ROSCillator:FREQuency[1|2]? [MIN|MAX]
This query returns the programmed frequency at specified port, in Hz or
minimum-maximum acceptable range limits.
Response format = <NR3>.
ROSCillator:MVOLtage?
This query returns the Oscillator oven monitor voltage.
Response format = <NR3>, precision = 1 x 10
-1
V.