Chapter 5 Command Reference
Description of SCPI Commands
92 Operating and Programming Manual
STATus:OPERation[:EVENt]?
This query returns the contents of the OPERation event register, then clears the
event register.
The event register bits are determined by the settings of the corresponding rise,
fall, and condition registers.
Response format = <NR1>.
STATus:OPERation:CONDition?
This query returns the contents of the OPERation condition register, non
destructively.
Response format = <NR1>.
STATus:OPERation:ENABle <numeric_value>
This command sets the OPERation enable register to the specified value.
Meaningful values are in the range 0 to 32767.
STATus:OPERation:ENABle?
This query reports OPERation enable register contents, non-destructively.
Response format = <NR1>.
STATus:OPERation:NTRansition <numeric_value>
This command sets the negative transition filter (fall register) as specified.
Meaningful values are in the range 0 to 32767.
STATus:OPERation:NTRansition?
This query returns the contents of the fall register.
Response format = <NR1>.
STATus:OPERation:PTRansition <numeric_value>
This command sets the positive transition filter (rise register) as specified.
Meaningful values are in the range 0 to 32767.

Table of Contents

Other manuals for Symmetricom 5071A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Symmetricom 5071A and is the answer not in the manual?

Symmetricom 5071A Specifications

General IconGeneral
BrandSymmetricom
Model5071A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals