Chapter 5 Command Reference
Description of SCPI Commands
Operating and Programming Manual 97
SYSTem:KEY? [MIN|MAX]
This query returns the numeric code of the last keypress or minimum-maximum
acceptable range limits.
Does not reflect codes sent by the KEY command.
Response format = <NR1>
SYSTem:PRINt?
This quires prints a status report.
This query or command is identical to the front-panel information print command.
Output format = arbitrary ASCII (several lines of ASCII text).
:REMote
SYSTem:REMote OFF|0|ON|1
This command enables or disables REMOTE operation mode.
The function of this command is ON at *RST.
SYSTem:REMote?
This query returns the state of REMOTE mode.
Response format = <0 or 1>
Returns 1 if REMOTE mode is ON, 0 if REMOTE mode is OFF
SYSTem:SCONfiguration
This command saves the following instrument parameters in nonvolatile memory:
Serial port: baud rate, data bits, stop bits, and parity
LCD angle
Output port frequencies
Printer report level
The 5071A must be in REMote mode to execute this command.
The function of this command is not affected by *RST.
SYSTem:TIME <hours>,<minutes>,<seconds>
This command is equivalent to ―SOUR:PTIM:TIME‖.
The function of this command is not affected by *RST.
SYSTem:TIME? [MIN|MAX, MIN|MAX, MIN|MAX]
This query returns 24-hour time or minimum-maximum acceptable range limits.
Response format = <NR1, NR1, NR1>, precision = 1 Second.

Table of Contents

Other manuals for Symmetricom 5071A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Symmetricom 5071A and is the answer not in the manual?

Symmetricom 5071A Specifications

General IconGeneral
BrandSymmetricom
Model5071A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals