7. Instrument Features
116 Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4 2008-07
7.2 Instrument Specifications
The table below lists the technical specifications of the instrument:
Parameters Characteristics
Measurement Software controlled
Interface USB
Filter Handling
infinite
®
F200
External filter exchange
Filter Handling
infinite
®
M200
Monochromator-based wavelengths change – no filters
necessary
Microplates to be
Measured
From 6-well to 384-well plates
(SBS standard formats)
Plate Definition Via scanning software
Temperature Control From 5 °C above ambient up to 42 °C
Plate Shaking Linear and orbital shaking, amplitude selectable from 1
– 6 mm in 0.5 mm steps
Parameters Characteristics
Light Source High energy Xenon Flash lamp, life time: 10
8
flashes
Optics Fused Silica Lenses
Fluorescence Detector Standard: PMT
Spectrally enhanced: red-sensitive PMT
Luminescence Detector Low dark count photomultiplier tube
photon-counting electronics (Basic and Standard PMT)
Absorbance Detector Silicon photodiode
Power Supply Auto-sensing: 100 – 120 V/220 – 240 V, 50-60 Hz
Power Consumption 150VA
Physical
Outer Dimensions
Basic instrument
Width: 425 mm
Height: 253 mm
Depth: 457 mm
16.73 inches
9.96 inches
17.99 inches
Weight infinite
®
F200
14 kg
Weight infinite
®
M200
15.8 kg
Outer Dimensions
Pump Module
Width: 250 mm
Height: 155 mm
Depth: 156 mm
9.84 inches
6.10 inches
6.14 inches
Weight Pump Module 3.4 kg
Environmental
Ambient Temperature
Operation + 15 °C to + 30 °C + 59 °F to + 86°F
Non-operation - 20 °C to + 60 °C -4 °F to + 140 °F
Relative Humidity
Operation < 90 % non condensing
Over-voltage Category II
Pollution Degree 2
Usage General Laboratory Instrument
Noise Level < 60 dBA
Method of Disposal Electronic waste
(infectious waste)