10. Troubleshooting
2008-07 Instructions for Use for infiniteĀ® 200 No. 30017581 Rev. No. 1.4 155
10. Troubleshooting
Error # Error Text Description
1 Command is not valid Unspecific error in the Instrument - Computer communication
protocol. Please report this error to your local Tecan
customer support office.
2 Parameter out of range Unspecific error in the Instrument - Computer communication
protocol. Please report this error to your local Tecan
customer support office.
3 Wrong number of parameters Unspecific error in the Instrument - Computer communication
protocol. Please report this error to your local Tecan
customer support office.
4 Invalid parameter Unspecific error in the Instrument - Computer communication
protocol. Please report this error to your local Tecan
customer support office.
5 Invalid Parameter at pos Unspecific error in the Instrument - Computer communication
protocol. Please report this error to your local Tecan
customer support office.
6 [prefix] is missing Unspecific error in the Instrument - Computer communication
protocol. Please report this error to your local Tecan
customer support office.
7 RS485 Timeout at module
[module descr]
Unspecific internal communication error. Please report this
error to your local Tecan customer support office.
8 Invalid module number [Nr] Unspecific error in the Instrument - Computer communication
protocol. Please report this error to your local Tecan
customer support office.
9 Binary Transfer command:
[cmd] at module [n]
Unspecific internal communication error. Please report this
error to your local Tecan customer support office.
10 Error at command [cmd] at
module [n],
Unspecific internal communication error. Please report this
error to your local Tecan customer support office.
11 LID open Plate transport or filter slide lid were open during a
measurement or the instrument was used in very bright
environment (<< 500 LUX). Please check if the lid closes
completely or if the environment was too bright.
12 LUMI FIBER broken Hardware Failure Luminescence Module. Please report this
error to your local Tecan customer support office.
13 Z Motor out of Safety-Range Unspecific error in the Instrument - Computer communication
protocol. Please report this error to your local Tecan
customer support office.
14 Filter is not defined Unspecific error in the Instrument - Computer communication
protocol. Please report this error to your local Tecan
customer support office.
15 X drive init error Hardware Failure Plate Transport Module. Please report this
error to your local Tecan customer support office.
16 Y drive init error Hardware Failure Plate Transport Module. Please report this
error to your local Tecan customer support office.