2. General Description
38 Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4 2008-07
2.4 Software
The infinite
®
200 is delivered with the i-Control software, for operating the
instrument and includes an online-help file and a printed Instructions for Use. The
software is formatted as a self-extracting archive on CD-ROM.
For advanced data reduction and full regulatory compliance with CFR 2 part 11
guidelines, The Magellan software can be used to control the
infinite
®
200. (For
more information, contact your local Tecan representative).
2.4.1 i-Control and Injectors
When using the injector, two modes are available:
• Dispense: The dispense mode allows liquid to be dispensed plate-wise
into the selected wells
• Injection: This mode must be used in combination with a measurement
strip. The injection is performed in a well-wise mode.
Dispense Mode
The dispense settings can be adjusted via the software:
Dispense
Select Injector: Injector A and/or Injector B can be selected.
Speed: The injection speed is selectable from 100 – 300 µl/sec for each
injector.
Select ‘Refill speed’ from 100 – 300 µl/sec. for each injector or select
‘Refill Speed equal to Dispense Speed’.
Select refill mode ‘Standard’, if refill should be performed when syringe is
empty (multiple dispense steps are performed before refilling, refill occurs
after dispensing approx. 800 µl).
Select ‘Refill for every dispense’ if refill should be performed for every
dispense step.