8. Quality Control
142 Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4 2008-07
8.3.7 Absorbance Wavelength Accuracy
The wavelength accuracy defines the deviation of the set measurement
wavelengths from the nominal wavelength. This test is only valid for the
infinite
®
M200.
Measurement Parameters:
Parameter Setting
Reading Mode Absorbance Scan
Measurement Wavelength from/to 300 – 850 nm
Step Size 1 nm
Number of flashes 25
Settle Time 0
Plate Type MultiCheck plate
Material/Reagents:
MultiCheck plate
Calculation of Wavelength Accuracy:
Please refer to the data sheet in the instructions for use for your MultiCheck plate.
mt
MaxMaxAccuracyWavelength
=
Max
t
theoretical Maximum
Max
m
measured Maximum