8. Quality Control
128 Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4 2008-07
8.2 Specifications - Passed/Failed Criteria
Note
All specifications are subject to change without prior notification.
The following table gives an overview of the passed/failed criteria for the
specification test of the
infinite
®
200.
Specification Cuvette (M200 only)
Passed/Failed Criteria
Absorbance Accuracy 0 – 2 OD: ≤ ± 1 % + 10 mOD
2 – 3 OD: ≤ ± 2.5 %
Absorbance Baseline Flatness
(1 sigma)
< ± 10 mOD
Specification Microplate
Passed/Failed Criteria
Fluorescence Top Sensitivity < 20 pM Fluorescein
Fluorescence Top Uniformity < 3 % CV
Fluorescence Top Precision < 2 % CV
Fluorescence Bottom Sensitivity < 100 pM Fluorescein
Fluorescence Bottom Uniformity < 3 % CV
Fluorescence Bottom Precision < 2 % CV
Time Resolved Fluorescence
Sensitivity (F200)
< 3 pM
Time Resolved Fluorescence Precision
(F200)
< 2 % CV
FP Precision (F200 only) < 5 mP
Luminescence Sensitivity Glow Type < 3 fmol/well (Standard PMT)
Luminescence Sensitivity Glow Type < 20 fmol/well (Basic PMT)
Luminescence Sensitivity Flash Type < 80 amol/well
(Standard PMT)
Luminescence Sensitivity Flash Type < 500 amol/well (Basic PMT)
Absorbance Accuracy 0 – 2 OD: ≤ ± 1 % + 10 mOD
2 – 3 OD: ≤ ± 2.5 %
Absorbance Baseline Flatness
(1 sigma)
< ± 10 mOD
Absorbance Wavelength Accuracy
≤ ± 1.5 nm
λ > 315 nm;
≤ ± 0.8 nm
λ ≤ 315 nm