9. Cleaning and Maintenance
150 Instructions for Use for infiniteĀ® 200 No. 30017581 Rev. No. 1.4 2008-07
9.3.2 Disinfection Procedure
WARNING
THE DISINFECTION PROCEDURE SHOULD BE PERFORMED IN A
WELL-VENTILATED ROOM BY AUTHORIZED TRAINED PERSONNEL
WEARING DISPOSABLE POWDER-FREE GLOVES, PROTECTIVE
GLASSES AND PROTECTIVE CLOTHING.
If the laboratory has no specific disinfection procedure, the following procedure
should be used to disinfect the outside surfaces of the instrument:
1. Disconnect the instrument from the main power supply.
2. Disconnect the instrument from any accessories that are used.
3. Carefully wipe all outside surfaces of the instrument with a wad of cotton
wool soaked in the disinfecting solution.
4. Make certain, that the same disinfection procedure is performed with the
plate carrier.
5. Repeat the disinfection procedure on any accessories, which are also being
moved for returned.
6. After the disinfection procedure has been performed, make certain that the
disinfection certificate is completed.
See 9.4 Disinfection Certificate for an example of the
disinfection certificate,
which must be completed before the instrument is returned to the service center
for service or repair.