8. Quality Control
2008-07 Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4 143
8.3.8 Absorbance Baseline Flatness (infinite
®
M200)
Perform the following measurement to determine the baseline-flatness:
Measurement Parameters:
Parameter Setting
Reading Mode Absorbance
Measurement Wavelength from/to 300 - 700 nm
Number of flashes 25
Settle Time 0
Plate Type GRE96ft
Part of the Plate A1
Plate Layout:
No plate is necessary for measurement – the plate carrier has to be empty for this
measurement.
Material/Reagents:
No material or reagents necessary for this test.
Calculation of Baseline Flatness:
Calculate the standard deviation.