2. General Description
2008-07 Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4 39
Using the Dispense Strip:
Plate
Select an appropriate plate type
Part of the plate
Optional;
Select the wells to be dispensed
Dispense
Set up the dispense parameters.
If both injectors are selected, all wells are first dispensed with
injector A and then with injector B.
The dispense strip does not require an additional measurement
strip.
Dispense volume
The injection volume depends on the microplate type. The plate
definition files include a so-called working volume. This working
volume defines the maximum volume to be dispensed into the
selected microplate. Therefore, always make sure that the
selected plate definition file contains the correct setting for the
working volume. The maximum dispense volume is
800 µl/dispense strip. If volumes greater than 800 µl are to be
dispensed (e.g. into 6-well plates), more than one dispense strip
has to be used.