Table 7-31
Miscellaneous Commands
Description
Command only. The INlt command causes the oscilloscope revert to the power-
on default states for the acquisition modes. The 2230 menu system is also
initialized.
With LONg
ON,
replies to queries are reported with the full command words.
With LONg
OFF,
replies use the short form of the command words. The short
form characters
are
those that appear
in
capitol letters
in
these command tables
and
are the minimum characters accepted
as
valid for commands. The power-
on
and
default states
of
LONg are
ON.
The LONg? query returns its state,
ON
or
OFF.
Returns the oscilloscope identification string
in
the form:
ID
TEK/2230,V81.1,VERS:09;
The instrument type
and
version numbers will
be
reported as appropriate for the
instrument queried.
Returns a list
of
all
the valid command headers available
in
the instrument
queried. All the valid characters
of
the commands are returned; the short form
of
the commands
(LONG
OFF)
are
in
capital letters.
Returns
an
ASCII string of headers
and
arguments reflecting the present states
of the controls
and
modes that may
be
set via the communications interface.
The query-only settings are not returned. The string returned by the SET? query
may
be
sent as a command message to the oscilloscope to recreate those
settings. The state of the LONg command affects the length of the reply.
NOTE
To
comply to Codes and Formats, a header is
not
sent back with the
settings string.
--