Appendix
A-2230
Operators
HORIZONTAL
Equipment Required (see Table A-1):
Calibration Generator (Item
1)
Leveled
Sine-Wave Generator (Item
2)
Time-Mark Generator (Item 3)
INITIAL CONTROL SETTINGS
Vertical
Channel
1 POSITION
VERTICAL MODE
X-Y
BW
LIMIT
CH
1 VOL TS/DIV
CH
1 VOL TS/DIV Variable
Channel 1 AC-GND-DC
Horizontal
POSITION
HORIZONTAL MODE
A SEC/DIV
SEC/DIV Variable
X10
Magnifier
B DELAY TIME POSITION
B TRIGGER
SLOPE
LEVEL
A TRIGGER
VAR
HOLDOFF
Mode
SLOPE
LEVEL
HF
REJECT
A&BINT
A SOURCE
A
EXT
COUPLING
Storage
Midrange
CH
1
Off (button out)
Off (button out)
0.5 V
CAL detent
DC
Midrange
A
0.05
µ.S
CAL detent
Off (knob
in)
Fully
counterclockwise
OUT
Fully
clockwise
NORM
P-P
AUTO
OUT
Midrange
OFF
VERT MODE
INT
DC+10
STORE/NON STORE
NON
STORE (button out)
SAVE/CONTINUE CONTINUE (button out)
PRETRIG/POST TRIG POST TRIG (button out)
ROLL/SCAN
SCAN
(button out)
1 K/4K
4K
(button out)
POSITION CURS/ POSITION CURS
SELECT WAVEFORM (button
in)
WAVEFORM REFERENCE/ WAVEFORM
REFERENCE
MENU
SELECT (button
in)
A-12
50 Q
BNC
Cable (Item
7)
50
Q
BNC
Termination (Item 10)
PROCEDURE STEPS
1. Check Timing Accuracy and Linearity
a.
Connect the time-mark generator output via a
50
Q
cable
and
a
50
Q termination to the
CH
1
OR
X input
connector.
b.
Select 50
ns
time markers from the time-marker
generator.
c.
Use the
Channel
1 POSITION control to center the
display vertically. Adjust the A TRIGGER LEVEL control
for a stable, triggered display.
d.
Use
the Horizontal POSITION control to align the
2nd
time marker with the 2nd vertical graticule
line.
e.
CHECK-
Timing accuracy
is
within
2%
(0.16 divi-
sion at the 10th vertical graticule
line),
and
linearity
is
within
5%
(0.1
division over
any
2 of the center 8
divisions). For checking the timing accuracy of the A
SEC/DIV switch settings from 50
ms
to 0.5
s,
watch the
time marker tips only at the
2nd
and
10th vertical graticule
lines while adjusting the Horizontal POSITION control.
f.
Repeat parts c through e for the remaining A
SEC/DIV
and
time-mark generator setting combinations
shown
in
Table
A-5
under the "Normal (X1)" column.
g.
Set:
A SEC/DIV
0.05
µ,S
X1
O Magnifier
On
(knob out)
h.
Select 10
ns
time markers from the time-mark
generator.
--