6-9
Troubleshooting
BPM PROBEBPM PROBE
MessageMessage Meaning and Corrective ActionMeaning and Corrective Action
(Arterial or Venous)
BPM-[SN] error
B001
B002
B004
B006
B007
B008
B010
B102
A problem detected with the BPM Probe electronics.
It may be possible to reset the BPM Probe by
disconnecting and reconnecting it to the Core.
If the issue persists, use a replacement BPM Probe
and contact Terumo Cardiovascular Technical
Support.
(Arterial or Venous)
BPM-[SN] SRS failure
B100
This failure indicates a probe has failed its reference
sensor test.
Ensure the BPM optical head is clean, undamaged
and not placed directly next to a light source. To
repeat the test, disconnect and reconnect the
BPM Probe to the Core. If the issue persists, use
a replacement BPM Probe and contact Terumo
Cardiovascular Technical Support.
(Arterial or Venous)
BPM-[SN]
Temperature Range Error
B101
The BPM Probe temperature measurement is out of
system range.
Conrm the BPM Probe is being used within the
specied environmental conditions in Appendix B.
Disconnect and reconnect the BPM Probe to the
Core. If the error persists, use a replacement BPM
Probe and contact Terumo Cardiovascular Technical
Support.
(Arterial or Venous)
BPM-[SN]
Sensor Intensity Error
B103
The BPM Probe sensor intensity measurement is out
of system range.
Conrm secure placement of the Shunt Sensor
in the BPM optical head. If the error persists, use
a replacement BPM Probe and contact Terumo
Cardiovascular Technical Support.
(Arterial or Venous)
BPM-[SN] error
B104
The calibration data stored on the Probe has been
corrupted.
Use a replacement BPM Probe and contact Terumo
Cardiovascular Technical Support.