6-9
Troubleshooting
BPM PROBEBPM PROBE
MessageMessage Meaning and Corrective ActionMeaning and Corrective Action
(Arterial or Venous)
BPM-[SN] error
B001
B002
B004
B006
B007
B008
B010
B102
A problem detected with the BPM Probe electronics.
It may be possible to reset the BPM Probe by
disconnecting and reconnecting it to the Core.
If the issue persists, use a replacement BPM Probe
and contact Terumo Cardiovascular Technical
Support.
(Arterial or Venous)
BPM-[SN] SRS failure
B100
This failure indicates a probe has failed its reference
sensor test.
Ensure the BPM optical head is clean, undamaged
and not placed directly next to a light source. To
repeat the test, disconnect and reconnect the
BPM Probe to the Core. If the issue persists, use
a replacement BPM Probe and contact Terumo
Cardiovascular Technical Support.
(Arterial or Venous)
BPM-[SN]
Temperature Range Error
B101
The BPM Probe temperature measurement is out of
system range.
Conrm the BPM Probe is being used within the
specied environmental conditions in Appendix B.
Disconnect and reconnect the BPM Probe to the
Core. If the error persists, use a replacement BPM
Probe and contact Terumo Cardiovascular Technical
Support.
(Arterial or Venous)
BPM-[SN]
Sensor Intensity Error
B103
The BPM Probe sensor intensity measurement is out
of system range.
Conrm secure placement of the Shunt Sensor
in the BPM optical head. If the error persists, use
a replacement BPM Probe and contact Terumo
Cardiovascular Technical Support.
(Arterial or Venous)
BPM-[SN] error
B104
The calibration data stored on the Probe has been
corrupted.
Use a replacement BPM Probe and contact Terumo
Cardiovascular Technical Support.