B-9
System Specifications
Electromagnetic Emissions Guidelines and Compliance Electromagnetic Emissions Guidelines and Compliance
The CDI OneView System is intended for use in the electromagnetic environment
specied in the following tables. The user of the system must ensure it is used in such
an environment.
Guidance and Manufacturer’s Declaration: Electromagnetic EmissionGuidance and Manufacturer’s Declaration: Electromagnetic Emission
Emission Test Compliance Electromagnetic Environment Guidance
Radiated
EMISSIONS
CISPR 11:2015
+A1:2016+A2:2019
Group 1/Class A
The CDI OneView System does not
generate any intentional RF. Therefore its
RF emissions are very low and unlikely to
cause interference in nearby electronic
equipment.
Conducted
EMISSIONS
CISPR 11:2015
+A1:2016+A2:2019
Group 1/Class A
The CDI OneView System is suitable
for use in all establishments other than
domestic and those directly connected to
public low-voltage power supply networks
that supply buildings used for domestic
purposes.
Harmonic current
EMISSIONS
IEC 61000-3-2
Class A
Voltage changes
+ uctuations
and icker
EMISSIONS
IEC 61000-3-3
Guidance and Manufacturer’s Declaration: Electromagnetic Immunity
Immunity Test IEC 60601-1-2
Test Level
Compliance
Level
Electromagnetic Environment
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floor should be wood, concrete
or ceramic tile. If oors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV (100
kHz repetition
frequency)
AC Mains
± 2 kV (100
kHz repetition
frequency)
AC Mains
Mains quality should be that of
a typical commercial or hospital
environment.
Always use a hospital-grade
power cord with the following
specications:
Length: 3m
Ampacity: 10A
Termination: IEC 60320-C13