B-11
System Specifications
Guidance and Manufacturers Declaration: Electromagnetic Immunity
Immunity Test IEC 60601-1-2
Test Level
Compliance
Level
Electromagnetic Environment
Guidance
Conducted RF
IE 61000-4-6
3 V
rms
150kHz to 80MHz
Outside ISM
bands
6 V
rms
ISM bands
between 150kHz
and 80MHz
80% AM at 1 kHz
3 V
rms
150kHz to 80MHz
Outside ISM
bands
6 V
rms
ISM bands
between 150kHz
and 80MHz
80% AM at 1 kHz
Portable and mobile RF
communications equipment
should be used no closer to
any part of the CDI OneView
System, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1.2 √P
150 kHz to 80 MHz
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from xed RF
transmitters, as determined by an
electromagnetic site survey
a
,
should be less than the
compliance level in each
frequency range
b
.
Interference may
occur in the vicinity of
equipment marked with
the following symbol: