4-16
Calibration and Circuit Setup
Attach Flow SensorAttach Flow Sensor
Follow the steps below to attach a Flow Sensor.
1. Open the Flow Sensor latch.
2. Seat the tubing in the sensor. Ensure the arrow on the sensor points in the
direction of the circuit ow.
3. Close the sensor latch and ensure it is securely sealed.
4. Wrap a color-coded identier around the Flow Sensor cable.
Note: Flow Sensors do not need to be calibrated.
Caution
For best accuracy, position the 1/4” Flow Sensor at least 3 inches
from any component in or attached to the extracorporeal circuit.
For best accuracy, position the 3/8” Flow Sensor at least 4 inches
from any component in or attached to the extracorporeal circuit.
321