4-16
Calibration and Circuit Setup
Attach Flow SensorAttach Flow Sensor
Follow the steps below to attach a Flow Sensor.
1. Open the Flow Sensor latch.
2. Seat the tubing in the sensor. Ensure the arrow on the sensor points in the
direction of the circuit ow.
3. Close the sensor latch and ensure it is securely sealed.
4. Wrap a color-coded identier around the Flow Sensor cable.
Note: Flow Sensors do not need to be calibrated.
Caution
For best accuracy, position the 1/4” Flow Sensor at least 3 inches
from any component in or attached to the extracorporeal circuit.
For best accuracy, position the 3/8” Flow Sensor at least 4 inches
from any component in or attached to the extracorporeal circuit.