6-2
Troubleshooting
STARTUPSTARTUP
MessageMessage Meaning and Corrective ActionMeaning and Corrective Action
System will not power on Ensure the Core is securely plugged into an AC
power outlet by conrming the white indicator
lights on both the Core and Display are illuminated.
Note: System will not power on while on battery
power.
If connected to AC power and the Core indicator
light is not illuminated, check the fuses located by
the AC power port on the Core. If a fuse is blown,
replace it with a 3.15 ampere (5mm x 20mm, 250V)
slow-acting fuse.
If the Display indicator light is not illuminated,
conrm the Display cable is securely connected to
both the Core and Display. If still not illuminated,
try a replacement Display cable.
If the issue persists, contact Terumo Cardiovascular
Technical Support.
Battery error
C504
A failure was detected in the battery during system
startup. The system can continue to be used while
connected to AC power.
Contact Terumo Cardiovascular Technical Support.
System error
C505
A failure was detected in the Core electronics.
Contact Terumo Cardiovascular Technical Support.
SYSTEMSYSTEM
MessageMessage Meaning and Corrective ActionMeaning and Corrective Action
Technical Alarm
Battery low
Shutdown in less
than 5 minutes
The system will alarm when operating on battery with
less than 5 minutes of charge remaining.
Connect to AC power.
Note: Connecting to AC power will deactivate the
alarm.