6-6
Troubleshooting
GAS CALIBRATIONGAS CALIBRATION
MessageMessage Meaning and Corrective ActionMeaning and Corrective Action
(Arterial or Venous) BPM
calibration failure
(pH, pCO
2
, pO
2
, K
+
intensity/slope)"
Shunt Sensor(s) have failed the calibration range
check to ensure proper performance per parameter.
This includes one or more of the following: arterial
pH, pCO
2
, pO
2
, K
+
and/or venous pH, pCO
2
, pO
2
, K
+
.
Retry gas calibration with new Shunt Sensor(s).
It is not recommended to repeat gas calibration
using the same Shunt Sensor. Buffer solution
in the sensor is lost during gas calibration, and
insufcient buffer solution may result in an
inaccurate gas calibration.
Ensure the sensor(s) are fully and properly engaged
with the BPM(s). Ensure proper placement of
the BPM(s) in the Calibrator. If the issue persists,
contact Terumo Cardiovascular Technical Support.
Note: if you are calibrating both Arterial and
Venous BPM assemblies and only one side fails, the
passing side will retain its new calibration data.
Calibrator Error
G006
G007
G010
G001
G008
G011
A problem was detected in the Calibrator
electronics.
It may be possible to reset the system by
disconnecting and reconnecting to the Core.
If the issue persists, contact Terumo Cardiovascular
Technical Support.
Barometric pressure
outside operational range
G400
This indicates an ambient pressure measurement
outside the acceptable range.
Conrm the device is being used within the
environmental specications listed in Appendix B.
If the error persists, a hardware failure is suspected.
Contact Terumo Cardiovascular Technical Support.
Gas (1 or 2) pressure
sensor error
G401
G402
The system is detecting gas bottle pressure out of
range.
Try replacing the corresponding gas bottle. If the
error persists, a hardware failure is suspected.
Contact Terumo Cardiovascular Technical Support.