7 Parameterizing, adjusting and analyzing software (P2A software)
121
3. Click on [OK] or [Cancel] to close the dialogue.
- The analog output returns to Measuring Mode again.
7.3.3.5. Displaying min./max. values
The transmitter saves the minimum or maximum value for each
channel (measured since the last voltage supply or since the last
manual reset).
✓ The required instrument file is marked.
1. Click on [Test/analyze transmitter].
- The Properties of <Instrument type> <Serial number>
dialogue is opened with the Test/analyze transmitter register.
2. Mark Min./max. values.
Field/button Explanation
Min./max. values View the min./max. values of each channel.
Only the values within the measuring range
are shown.
Channel Min./max. of channel 1/2/3 (optional)
Value Min. or max. value, 1 decimal place.
Unit Unit selected in
.
3. Reset Min./max. values.
4. Click on [Reset min./max. values].