7 Parameterizing, adjusting and analyzing software (P2A software)
121
3. Click on [OK] or [Cancel] to close the dialogue.
- The analog output returns to Measuring Mode again.
7.3.3.5. Displaying min./max. values
The transmitter saves the minimum or maximum value for each
channel (measured since the last voltage supply or since the last
manual reset).
The required instrument file is marked.
1. Click on [Test/analyze transmitter].
- The Properties of <Instrument type> <Serial number>
dialogue is opened with the Test/analyze transmitter register.
2. Mark Min./max. values.
Field/button Explanation
Min./max. values View the min./max. values of each channel.
Only the values within the measuring range
are shown.
Channel Min./max. of channel 1/2/3 (optional)
Value Min. or max. value, 1 decimal place.
Unit Unit selected in
Unit/analog output
.
3. Reset Min./max. values.
4. Click on [Reset min./max. values].