7 Parameterizing, adjusting and analyzing software (P2A software)
122
5. Confirm control query to perform the reset.
- The values are reset to the factory settings.
6. Click on [OK] or [Cancel] to close the dialogue.
7.3.4. Adjusting the transmitter
This function is used to adjust an attached instrument. The
following adjustments may be carried out using the software:
• 1-point adjustment (offset)
• 2-point adjustment (upper and lower adjustment point)
• Analog adjustment (entry via assistant/wizard)
• n-point adjustment (entry via assistant/wizard)
Also see 4.3.2.6. volume 1, see Adjusting the instrument,.
7.3.4.1. 1-point adjustment
The testo 400/650 with precision humidity probe (order no.
0636 9741) is recommended as the reference measuring
instrument for 1-point adjustment (offset) (see, 4.3.2.8.,
volume 1, Adjusting testo 6383 using testo handheld
instrument).
1. Expose the reference measuring instrument and the instrument
to be adjusted to the same constant conditions and wait for
equalization period to lapse.
2. Mark the instrument file of the connected instrument.
3. Click on [Adjusting the transmitter].
- The Properties of <Instrument type> <Serial number>
dialogue is opened with the Adjusting the transmitter register.
4. Enter reference value and click on [Carry out 1-point
adjustment].