6 testo 6610 probes
84
6.2.4. testo 6614 heated cable probe
The testo 6614 heatable probe is suitable for use in high-humidity
processes in which there is the possibility of condensation for the
probe.
Also see 4.3.2.12. volume 1, High-humidity adjustment for
testo 6614, for the functional principle of the testo 6614.
At a glance
Connector
Key
Probe cable
Probe shaft
Filter (including:
humidity and
temperature sensor)
Temperature probe
Cable for
temperature probe
Condensation
protector
(0554 0166)
In the event of overpressures, the probe may become a
projectile.
> Mount the probe pressure-
tight (see pressure resistance under
Technical data)
Application
• Monitoring and regulating high-humidity processes, e.g. drying
(ceramics, tobacco, wood, food) and maturing (cheese, fruit).
• At flow velocities > 1 m/s, an additional measuring uncertainty,
not higher than +1.5 % RH, is to be expected. To prevent this
error and ensure the highest measuring accuracy: Use
condensation protector 0554 0166.