Message Display Cause Remedying of fault
01116 High adjustment
temperature
The ambient
temperature is too high
during the pressure
adjustment
Take necessary
measures to lower
ambient temperature,
e.g. through venting.
03000 Heat function
defective
For testo 6614 probe
only: Heat function
defective
Contact Testo Service
4.5.4. Handling alarm messages
Shown on the
4
Can be used for
5
Additional
6
New limit value x
Scaling changed x
Pressure too high x x
Delta p adjustment x
Alarm 1 x
Alarm 2 x
Alarm 3 x
Alarm 4 x
Transmitter reset x
Reset Min/Max
Transmitter refresh x
User Setting Change
Analog adjustment x
1-point adjustment x
2-point adjustment
x
4
If multiple messages/alarms are activated at the same time, only the last message/alarm is shown. If
this is cancelled, the other messages that are still active are no longer displayed.
5
The message can be assigned the collective alarm function, which means that the collective alarm is
activated as soon as at least one of the messages assigned to it is activated. The collective alarm can
be assigned to each of the 4 optional relays. The collective alarm is then always the same, as it can
only be defined once.
6
The message is shown upon the occurrence of the event causing the message as well as when
closing. Two entries appear in the history in the P2A software: Message text_start and Message
text_end.