6 testo 6610 probes
74
6.2. Product description
6.2.1. Overview of probe and filter types
6.2.1.1. Probe version
A detailed description of the probe versions can be found from see
6.2.2. Integrated humidity probe, onwards.
The following probe versions are available for the testo 6383
transmitter:
Version Article no. Characteristic
Integrated
humidity
probe
KMAT option
D04
(0636 6610)
Wall probe version; accuracy to ±1 % RH;
temperature range -20 to +70 °C/-4 to +158 °F,
plug-on sensor
testo 6612 0555 6610-L12 Duct probe version; accuracy to ±1 % RH;
temperature range -30 to +150 °C/
-22 to +302 °F, sensor soldered
testo 6613 0555 6610-L13 Cable probe version; accuracy to ±1 % RH;
temperature range -70 ... +180 °C /
-94...+356 °F, sensor soldered
testo 6614 0555 6610-L14 Heated cable probe version; accuracy to ±1 %
RH; temperature range -40 to +180 °C/
-40 to 356 °F, sensor soldered
testo 6615 0555 6610-L15 Trace humidity cable
probe version; accuracy to
±6 K at -60 °Ctd; temperature range -40 to
+120 °C/-40 to +248 °F, sensor soldered
testo 6617 0555 6610-L17 Cable with cover electrode monitoring probe
version; accuracy to ±1.2 % RH; temperature
range -40 to +180 °C/
-40 to +356 °F, sensor soldered
6.2.1.2. Determining the accuracy/measuring uncertainty
The entries for the measuring uncertainty for the probe are
determined in accordance with GUM (Guide to the Expression of
Uncertainty in Measurement/DIN V ENV 13005).
All parts that make up the measuring uncertainty given by Testo
are listed below. When comparing the measuring
uncertainty/accuracy between manufacturers, which components
are included is to be taken into account. In many cases, not all
elements that contribute to measuring uncertainty are assessed, for
example if the error contribution of the production adjustment for
some manufacturers is shown separately or not at all.