EasyManua.ls Logo

TESTO 6383 - testo 6617 cable probe (self-monitoring)

TESTO 6383
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 testo 6610 probes
91
6.2.6. testo 6617 cable probe (self-monitoring)
The testo 6617 probe is used if spatial separation of the transmitter
and probe is required, particularly for media (gases, vapours) which
can jeopardize/damage the humidity sensor (the testo 6617 has a
self-monitoring and early-warning function for such applications).
At a glance
1
Connector
2
Key
3
Probe cable
4
Probe shaft
5
Filter (including: humidity
and temperature sensor)
WARNING
In the event of overpressures, the probe may become a
projectile.
> Mount the probe pressure-tight (see pressure resistance under
Technical data)
Application
Monitoring and regulating industrial humidity processes
involving corrosive media (apart from high-humidity processes)
with the exception of applications with HCL, HF and other acids
and acidifiers in larger concentrations (SO
2
, SO
3
, NO
2
)
Meeting the top accuracy requirements
Applications for which a sturdy metal housing is required
This probe does NOT have a longer life with exposure to
aggressive media than the testo 6613, for example.
However, early warning of damage to the sensor is given
so that machine failures can be avoided.