4.5.2. Warning messages
Warning messages show an early warning or a current malfunction
which may negatively impact measuring.
Message Display Cause Remedying of fault
00809 Pressure too
high**
The process pressure
exceeds the pressure
intended for the
transmitter
Remove the transmitter
from the process and
take any necessary
measures to lower the
pressure
00E00 T ambient high**
The ambient
temperature exceeds
the permissible
temperature for the
transmitter
Take necessary
measures to lower
ambient temperature,
e.g. through venting or
cooling
00E01 T ambient low** The ambient
temperature is below
the permissible
temperature for the
transmitter
Take necessary
measures to raise
ambient temperature,
e.g. through heating
00E04/00E
05
Supply voltage
low**
The supply voltage is
below the minimum
voltage required for the
transmitter
Take measures to
ensure sufficient
voltage supply
02822 T process high** The process
temperature exceeds
the temperature
designated for the
probe
the process and take
any necessary
measures to lower the
process temperature
02821 T process low** The process
temperature is below
the temperature
designated for the
probe
Take necessary
measures to raise
process temperature
0081C Alarm 1** Depending on the
parameterization of the
relays
Depending on the
parameterization of the
relays
0081D Alarm 2** Depending on the
parameterization of the
relays
Depending on the
parameterization of the
relays