4 Transmitter
54
9. With or continue to Analog Adj. Ch. 2or 3 (repeat steps 3
to 8).
10. Return to main menu Adjust with ESC.
11. Continue to main menu Reset with or or return to
Measuring Mode with ESC.
Performing the pressure adjustment
Please refer to 4.3.2.11. n-point adjustment (pressure).
1. In the Measuring Mode press SET, select main menu Adjust
with or and confirm selection with SET.
The pressure adjustment can be performed at 3 to 6
adjustment points.
2. Select Adjust Pressure with or and confirm with SET.
3. Use or to select Adj. Point 1.
4. Press SET. Read off pressure sensor display (e.g. 30.1 Pa) and
enter this value in the user menu. Do this by scrolling one digit
to the right using and increasing the value of digit by 1 using
. Confirm with SET or abort entry with ESC.
5. Repeat steps 1 to 4 for adjustment points 2 to 6.
6. Return to main menu Adjust with ESC.
7. Continue to main menu Reset with or or return to
Measuring Mode with ESC.
4.4.5.10. Editing Reset main menu
You can reset the factory settings for the following:
Instrument
Sensor/probe
Min./max. values
Resetting to the factory settings means resetting to the
order specification, i.e. the specific condition at the time of
supply to the customer.
1. In the Measuring Mode press SET, select Main Menu Reset
with or and confirm selection with SET.
- Reset Device to factory settings is displayed.
2. Select the setting to be reset using or and confirm
selection with SET.
- Reset Completed is displayed.
3. Return to the reset setting with ESC or SET and return to Main
Menu Reset with ESC.
4. Continue to Main Menu Channel 1 with or or return to
Measuring Mode with ESC.