7 Parameterizing, adjusting and analyzing software (P2A software)
123
5. Confirm confirmation request.
- The adjustment is carried out.
Field
Explanation
°C/°F
Selection of the unit; only for temperature
adjustment.
Current reading Reading in °C/°F or % RH.
Readings are updated every second.
Reference value Entry of the read-off value from the reference
measuring instrument.
Permissible entries:
max. 5 % RH deviation (sum of all 1-point
adjustments)
max. 2 K (°C) deviation (sum of all 1-point
adjustments)
> Click on [Set offset to zero] to reset a transmitted reference
value.
- The current reading is used again.
6. Click on [OK] or [Cancel] to close the dialogue.