4 Transmitter
62
higher-level control system. The values correspond to the "Namur"
industry standard.
Status
message in
the display
Display
value in
display
Class Analog output
0 to
20 mA
4 to 20
mA
1 V 5 V 10 V
No probe
signal
None Error 21 mA 21 mA 1.1 V 5.5 V 11 V
Wrong probe None Error 21 mA 21 mA 1.1 V 5.5 V 11 V
Values < 0
%RH
uuuuu Underrange 0 mA 3.8 mA 0 V 0 V 0 V
Condensation
ooooo Overrange 20.5 mA 20.5 mA 1.1 V 5.5 V 11 V
% RH short-
circuit
----- Error 21 mA 21 mA 1.1 V 5.5 V 11 V
% RH sensor
broken
----- Error 21 mA 21 mA 1.1 V 5.5 V 11 V
T short-circuit ----- Error 21 mA 21 mA 1.1 V 5.5 V 11 V
T sensor
broken
----- Error 21 mA 21 mA 1.1 V 5.5 V 11 V
T process low
uuuuu Underrange 0 mA 3.8 mA 0 V 0 V 0 V
T process
high
ooooo Overrange 20.5 mA 20.5 mA 1.1 V 5.5 V 11 V
Probe
disconnected
None Error 21 mA 3.8 mA 1.1 V 5.5 V 11 V
Watchdog
error
Previous
value
stops
Error 21 mA 3.8 mA 1.1 V 5.5 V 11 V
Value below
min. scale
Reading Underrange 0 mA 3.8 mA 0 V 0 V 0 V