6 testo 6610 probes
82
Probe shaft: Stainless steel
Line: Sheathed, FEP
Connector: ABS plastic
Accuracy (at 25 °C/77 °F)*
Humidity
± (1.0 % RH + 0.007 x reading) for 0 to 90 % RH
± (1.4 % RH + 0.007 x reading) for 90 to 100 % RH
0.02 % RH/K, dependent upon the process temperature
(with a deviation of 25 °C/+77 °F)
0.02 % RH/K, dependent upon the electronics temperature
(with a deviation of 25 °C/+77 °F)
Temperature
±0.15 °C (0.27 °F) with slope PT1000 class AA
* Refer to the charts below for the correlation between
temperature and accuracy.
Reproducibility
Better than ±0.2 % RH
Sensor
Response time without protective filter: t 90 max. 15 sec.
Probe dimensions
Diameter of probe shaft: 12 mm
L = approx. 120/200/300/500/800 mm
L A = 85/165/265/465/765 mm
A, 6.2.1.4. see Overview of probe and filter types.
Probe length incl. probe shaft and filter
1/2/5/10 m
Pressure resistance**
PN 10 (probe tip)
PN 1 (if end of probe/cable is involved in process)
** If installing probe under pressure, please use cutting ring screw
connection (order no. 0554 1795).
Measuring accuracy of testo 6613 cable probe
Humidity error, absolute |±% RH| as a factor of process humidity
(% RH)