6 testo 6610 probes
89
Measuring range
Dewpoint: -40 to + 30 °C
td
/-148 to +212 °F
td
Temperature: -40 to ±120 °C/-40 to +248 °F
(Temp. durability up to +180 °C/+356 °F)
Material
Probe shaft: Stainless steel
Line: Sheathed, FEP
Connector: ABS plastic
Accuracy (at 25 °C/77 °F)*
Dewpoint
± 1 K at 0 °C
td
/+32 °F
td
± 2 K at -40 °C
td
/-40 °F
td
± 4 K at -50 °C
td
/-58 °F
td
± 6 K at -60 °C
td
/-76 °F
td
Dependent on the process temperature (deviation of 25 °C /
77 °F)
± 0.1 Ktd/K at -40 °C to 25 °C/-40 °F to +77 °F
± 0.2 Ktd/K at 25 °C to 50 °C/77 °F to 122 °F
± 0.4 Ktd/K at 50 °C to 120 °C/122 °F to 248 °F
Temperature
±0.15 °C (0.27 °F), slope PT1000 class AA
* Refer to the charts below for the correlation between
temperature and accuracy.
Sensor
Response time without protective filter: t 90 max. 15 sec.
Probe dimensions
Diameter of probe shaft: 12 mm
L = approx. 200/500 mm
L A = 165/465 mm
A see Overview of probe and filter types, page 74.
Cable length incl. probe shaft and filter
1/2/5/10 m
Pressure resistance
PN 16 (probe tip)**
** If installing probe under pressure, please use cutting ring screw
connection (order no. 0554 1795).