6 testo 6610 probes
94
Accuracy (at 25 °C/77 °F)*
• Humidity
◦ ± (1.2 % RH + 0.007 x reading) for 0 to 90 % RH
◦ ± (1.6 % RH + 0.007 x reading) for 90 to 100 % RH
◦ 0.02 % RH/K, dependent upon the process temperature
(with a deviation of 25 °C/+77 °F)
◦ 0.02 % RH/K, dependent upon the electronics temperature
(with a deviation of 25 °C/+77 °F)
• Temperature
◦ ±0.15 °C (0.27 °F) with slope PT1000 class AA
* Refer to the charts below for the correlation between
temperature and accuracy.
Reproducibility
• Better than ±0.2 % RH
Sensor
• Response time without protective filter: t 90 max. 15 sec.
Probe dimensions
• Diameter of probe shaft: 12 mm
• L = approx. 200/500 mm
• L – A = 165/465 mm
• A see Overview of probe and filter types, page 74.
Probe length incl. probe shaft and filter
• 1/2/5/10 m
Pressure resistance
• PN 10 (probe tip)**
** If installing probe under pressure, please use cutting ring screw
connection (order no. 0554 1795).