■
1-4
- Be extremely careful not to take too much time or too many measurements, as
this can stress the patient.
- Constantly observe that both the instrument and patient are free of problems.
- If a problem with the instrument or the patient occurs, take appropriate action,
such as stopping the device, to ensure the patient’s safety.
- Do not allow the patient to touch the instrument.
- If any smoke, offensive odor, or abnormal sound occurs, turn off the
instrument immediately, disconnect the power plug from the outlet, and contact
our local distributor or Tomey Corporation.
■
Precautions after operation
- Do not place any container with liquid in it on the instrument. Any liquid
entering the instrument may cause electric shock or failure.
- Do not use organic solvents such as thinner, benzene, or acetone to clean the
instrument. Fire or electric shock may occur. (These solvents may also
corrode the resin or coating on the cover of the instrument.)
- Do not allow water or chemicals to splash on probes, except for the tips. Any
water or chemicals entering the instrument may cause an electric shock or
failure.
- Follow the specified procedures to return the operation switch, dial, etc. to
their original positions and turn the instrument off.
- Hold the plug when disconnecting the power plug from the outlet to avoid
placing excessive force on the cord. Pulling the cord may damage the inner
core wires, resulting in electric shock or fire.
- When disconnecting cords, do not apply too much force to them, for example,
do not hold and pull the cord.
- Refer to “5.6 Storing” for instructions on the storage of the instrument.
- Clean the instrument at the end of operation in preparation for the next use.
- Clean and neatly arrange the accessories and cables.