3-225
■
(2) BD Address
Enter the BD address of the connection place.
The BD address of the connection place is shown in “Bureau / BD Address” on the wireless
communication setting screen of the utility function on the connected device.
(3) Search button for BD address
Opens the BD address search screen.
<BD address setting examples>
The following describes the settings on each device when the BD address of each device to be
connected is as listed in Table 1.
Table 1: BD address examples of each device
BD address example
of this instrument
00A0962F9010 Connection place
AL-4000 measurement
unit BD address example
00A0962F9520
Table 2: Setting contents on “connection place setting screen” of each device
Settings of this instrument Settings of AL-4000 measurement unit to be
connected
Connection place
name
AL4000 MEAS2 Connection place name IOL_CAL
BD address 00A0962F9520 BD address 00A0962F9010
1) When starting connection with the settings as listed in Table 2 for this instrument, the BD
address of this instrument is automatically set at the connection place.
- Connection place name: Set the desired name. (Both devices)
- BD address: Set the BD address of the connected device. (This instrument only)
2) Select the device to be connected as the “Connection place” on the wireless
communication setting screen on this instrument (the device can be selected by (5) on
the wireless communication setting screen in this manual).
3) Select “Permit” for “Wireless Communication” on the wireless communication setting
screen on both devices.
(4) “Cancel” button
Cancels the entered data and returns to the top page of the wireless communication setting
screen.
(5) “Apply” button
Applies the entered data and returns to the top page of the wireless communication setting
screen.