39
: Field
-0.090%
-0.070%
[ LOW ]
-0.050%
0.00010
LIMIT
FUNC: Cp-D
COMP: OFF
ALARM OFF
AUX: OFF
MODE: %TOL
NOMINAL:
217.259nF
3
2nd:
LIST
<Limit Table>
[BIN]
2
1
0.090%
0.00014
CORR SETUP
0.070%
0.050%
[ HIGH ]
ToolsFile
Figure 4-14. Available Fields on the Limit Table Page
LOW
MODE
NOMINAL
<Limit Table>
p
(TH2816A)
[NUMERIC ENTRY]
PAGE
DEL--
%TOL
±TOL
NEXT
μ
ESC
n
p
COMP
ON
OFF
μ
ESC
n
[NUMERIC ENTRY]
SETUP
CORR
LIMIT
LIST
NEXT
LOAD
SAVE
File
ESC
NEXT
m
DEL--
PAGE
[NUMERIC ENTRY]
HIGH
NEXT
ESC
k
M
m
ESC
M
k
(TH2816A)
n
μ
p
AUX
ON
OFF
SWAP
FUNC
CLEAR
Tools
NEXT
ESC
ESC
OK NO
YES
NEXT
ESC
k
M
m
Figure 4-15. Available Softkeys on the Limit Table Page
4.3.4.1 Nominal Value for tolerance mode
When the tolerance mode is used as a limit mode for the primary parameter, the
nominal value must be set.
Front Panel Operation for Setting the Nominal Value
Perform the following steps to set the nominal value.
1) Move the cursor to the NOMINAL field.
2) Enter the nominal value using the numeric entry keys. When the numeric data is
entered, the unit softkeys (p , n , μ, m , k , M ) can be used instead of ENTER.