6
z Absolute deviation or percent deviation can be set as comparison limits.
TH2816B can output comparison results for sorting components into a maximum
of 10 bins. (BIN1 to BIN 9 and OUT). Also, devices whose primary parameter is
within limits, but whose secondary parameter measurement result is not within
limits, can be sorted into an AUXiliary BIN.
TH2816B provides several communication interfaces which make it easy to output
the measurement results to other equipments (for example computers), or build an
automatic test system.
z RS232C interface: RS232C interface make it easy to communicate with other
equipments. You can set the measurement functions and input the parameters
through the RS232C interface instead of front panel keyboard.
The RS232C interface can also be used as the control interface for the external
bias current source.
z IEEE488 interface (optional): The IEEE488 interface is an optional interface on
the TH2816B and can be used to build an automatic test system to completely
characterize new components and materials, and to fully automate production
line testing.
z HANDLER interface(optional): By using the handler interface, the TH2816B can
easily be combined with a component handler, and a system controller to fully
automate component testing, sorting, and quality control data processing.
TH2816B’s file function can save the setup parameters, limit parameters and list
sweep parameters as a file in the internal nonvolatile memory, and the No. 0 file will
be automatically reloaded when TH2816B is powered on.
1.2 Operation conditions
1.2.1 Power requirements
TH2816B requires the following power source:
Voltage: 198 to 242 Vac
Frequency: 47 to 66 Hz
Power: 50 VA maximum
1.2.2 Operation environment
TH2816B must be operated under within the following environment conditions.
Temperature0 to 40
Humidity: less then 90% RH
1.2.3 Warm-up time
Allow TH2816B to warm up a minimum of 20 minutes before starting tests.
1.2.4 Notes for usage
z Please do not operate the instrument in the places where is vibrative, dusty,
under direct sunlight, or where there is corrosive air.
z Although the instrument has been specially designed for reducing the noise
cased by ac power, a place with low noise is still recommended. If this cannot be
arranged, please make sure to use power filter for the instrument.
z Please store the instrument in the place where temperature is between 5 and
40, humidity is less then 85% RH. If the instrument will not be put in use for a
time, please have it properly packed with its original box or a similar box for
storing.
z TH2816B has the cooling fan on the rear panel and cooling holes on both sides.
High temperature inside will decrease the measurement accuracy, so sufficient
space must be kept around the TH2816B to avoid obstructing the air flow of the