73
OPTIONAL OPERATIONS
IOL MODE IN TONO/PACHO
SETTING THE IOL MODE
1 Check the measurement screen; If displays beneath AUTO/MANUAL display,
IOL mode is set.
2 If no is displayed, tap the button to change to IOL mode.
SETTING THE IOL CAMERA FOCUS
1 If more than 2 alignment dots appear, adjust focusing point using the
button in "Camera Focus" so that alignment dot becomes one.
2 Tap the button and IOL mode setting screen disappears.
NOTE
Alignment may not be performed normally with IOL inserted eye. If it occurs,
carry out measurement in IOL mode.
NOTE
The value sets at "Camera Focus" remains in the "IOL Camera Focus" in
"TONO/PACHO" of the "Initial".
IOL
PatientID
PatientID
/1
/1
/1
/1
/1
/1
/1
/1
1
1
T
OPCON
T
ARO
U
TOPCONTAROU
MODE
T/P
Pleasetouch
pupilonthescreen.
PatientID
PatientID
/1
/1
/1
/1
/1
/1
/1
/1
1
1
T
OPCON
T
ARO
U
TOPCONTAROU
MODE
T/P
Pleasetouch
pupilonthescreen.
IOL mode setting screen IOL mode display
Focus forward/backward
PatientID
PatientID
/1
/1
/1
/1
/1
/1
/1
/1
1
1
T
OPCON
T
ARO
U
TOPCONTAROU
MODE
T/P
Pleasetouch
pupilonthescreen.
Focus forward button
Focus backward button
Exit
/1
/1
/1
/1
1
1
Exit button