MAINTENANCE
FUSE CHANGE
1 Make sure the power is off and the power cable is unplugged.
2 Tilt the body slowly so that the switch comes up and the power inlet at the bottom
can be seen.
3 Press the fuse holder with a screwdriver and turn it counter-clockwise. The fuse holder can be
taken out.
4 Replace the fuse with a spare one.
5 Press the fuse holder until it clicks.
The fuse holder is set.
WARNING
To avoid electric shock during fuse change, be sure to unplug the
power cable before removing the fuse lid.
Also, do not plug the power cable while the fuse lid is removed.
Always use the attached fuse (T 2AL 250V). Using any other
type may cause malfunction and/or fire.