OPTIONAL OPERATIONS
MANUAL MODE IN TONO/PACHO
SETTING THE MANUAL MODE
1 Check that the MEASUREMENT screen is on. If the button is "M," the mode is
Manual mode.
2 If "A" (Auto mode) is displayed, tap it and change to "M".
SETTING THE MEASURING RANGE
In this instrument, the measuring range can be switched in 2 steps between "1-30" and "1-60."
Normally, "1-30" is used, but if the patient's ocular pressure is high, switch it to "1-60." The default
setting is "1-30" upon power on.
1 Check the measurement screen.
2 Tap the button of the control panel and set the measuring range.
NOTE
Adjust the height of the instrument table so that the patient can sit com-
fortably. Otherwise, correct measurement values may not be obtained.
Do not perform measurement if the patient holds their breath or is ner-
vous. Otherwise, correct measurement values may not be obtained.
If the machine is moved before measurement values are displayed, it may
cause incorrect measurement results.
The beep function for urging cautions is provided so that a finger or a
hand is not pinched between the reverse side of forehead rest, a measur-
ing head and an ocular pressure measurement window.
Auto/Manual
PatientID
PatientID








1
1
T
OPCON
T
ARO
U
TOPCONTAROU
Focusand
Touchameasurepoint
T/P
MANUAL
MODE
30/60
PatientID
PatientID








1
1
T
OPCON
T
ARO
U
TOPCONTAROU
Focusand
Touchameasurepoint
T/P
MANUAL
MODE