48
BASIC OPERATIONS
BASIC OPERATIONS
MEASUREMENT IN REF/KRTTONO/PACHO CONTINUOUS MEASUREMENT MODE
DISPLAYING MEASUREMENT VALUES
With regard to measurement values, for REF, KRT, TONO and PACHO, data of the latest measure-
ment (only for TONO/PACHO, latest 3 measurements) are displayed on the control panel.
Figures only: Measurement was done correctly.
[ ] figures: When the reliability of measurement is low. (only TONO)
ERROR: Measurement was not done correctly.
OVER: When the measurement range is exceeded. (only TONO)
NOTE
In TONO average value display, low-reliability numerical data with [ ] are
not added to average value calculation.
However, if all measurement data are numerical data with [ ], average
value calculation is done using these data.
For explanation of the messages on the control panel screen, refer to
"MESSAGE LIST" on page 99.
In the data printout, manual measurement values will have M marks
beside them. (only for TONO/PACHO)
When the "Auto print" is set to the "OFF" in the "Common" of the "Initial",
print out measurement results by tapping the button, as neces-
sary.
Print out