Cleaning the source hexapole assembly 5-51
7. Inspect the hexapole assembly for persisting contamination, If
contamination is present, do as follows:
a. Use the wash-bottle containing methanol to rinse the source
hexapole assembly over the large beaker.
b. Blow-dry the hexapole assembly with inert, oil-free gas.
8. Use the small flat-blade screwdriver to ensure that the hexapole
assembly screws are tight.