Removing the IonSABRE APCI probe 4-7
Removing the IonSABRE APCI probe
Required material: Chemical-resistant, powder-free gloves
To remove the IonSABRE APCI probe
1. Prepare the instrument for working on the source (see “Preparing the
instrument for work performed on the source” on page 5-7).
2. Open the instrument’s access door.
3. Disconnect the diverter valve tubing from the IonSABRE APCI probe.
4. Disconnect the IonSABRE APCI probe’s electrical lead from the
instrument’s probe connection.
5. Disconnect the IonSABRE APCI probe’s PTFE tube from the nebulizer
gas connection.
6. Undo the 2 thumbscrews securing the probe to the probe adjuster
assembly.
7. Carefully remove the probe from the probe adjuster assembly.
8. Close the instrument’s access door.
Warning: The ACQUITY UPLC system connections, IonSABRE
APCI probe, and source can be contaminated with biohazardous
and/or toxic materials. Always wear chemical-resistant,
powder-free gloves while performing this procedure.
Warning: To avoid electric shock, ensure that the instrument is
suitably prepared before commencing this procedure.
Warning: The source can be hot. To avoid burn injuries, take
great care while working with the instrument’s access door open.