Replacing the source assembly seals 5-83
Removing the seals from the source enclosure and probe
adjuster assembly
Required materials:
Chemical-resistant, powder-free gloves
Flat-blade screwdriver
O-ring removal kit
To remove the seals from the source enclosure and probe adjuster
assembly
1. Use the O-ring removal kit to carefully remove the probe adjuster
assembly probe seal from the probe adjuster assembly (see “Removing
O-rings and seals” on page 5-10).
Warning: The source components can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
TP02632
Probe adjuster assembly probe seal