EasyManua.ls Logo

Waters TQ Detector - Replacing the ESI Probe Tip

Default Icon
235 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-54 Maintenance Procedures
Replacing the ESI probe tip
Replace the ESI probe tip if a blockage occurs in the internal metal sheathing
through which the stainless steel capillary passes or if the probe tip threads
are damaged.
Required materials:
Chemical-resistant, powder-free gloves
6-mm (¼-inch) wrench
To replace the ESI probe tip
1. Remove the probe from the source (see also “Removing the ESI probe” on
page 3-10).
2. Use the 6-mm (¼-inch) wrench to unscrew and remove the probe tip.
3. Dispose of the probe tip in accordance with local environmental
regulations.
4. Use the 6-mm (¼-inch) wrench to fit and tighten the new probe tip to the
probe.
Warning: The probe and source components can be contaminated
with biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning: The probe and source can be hot. To avoid burn injuries, take
great care while performing this procedure.
Warning: The ESI probe tip is sharp. To avoid puncture wounds,
handle the probe with care.
Warning: The probe tip can be contaminated with
biohazardous and/or toxic materials. Ensure that it is
correctly disposed of according to local environmental
regulations.

Table of Contents

Other manuals for Waters TQ Detector

Related product manuals