Installing the IonSABRE APCI probe 4-3
Installing the IonSABRE APCI probe
Probe compatibility warning label:
Required material: Chemical-resistant, powder-free gloves
Warning: On instruments bearing the probe compatibility
warning label (below), always use an APCI probe whose design
incorporates a drain spout. In the event of a leak at the capillary
union, using an older APCI probe model, whose design does not
incorporate a drain spout, can lead to unmanaged solvent
spillage and an associated risk of ignition.